Site icon New

MCQ – 90316

Question

Subject/object unsharpness can result from all of the following except when
A.

object shape does not coincide with the shape of x-ray beam

B.

object plane is not parallel with x-ray tube and/or IR

C.

anatomic object(s) of interest is/are in the path of the CR

D.

anatomic object(s) of interest is/are at a distance from the IR

Show Answer

Correct Answer � C

Explanation

Answer (C)

  • A certain amount of object unsharpness is an inherent part of every radiographic image because of the position and shape of anatomic structures within the body. Structures within the three dimensional human body lie in different planes. In addition, the three-dimensional shape of solid anatomic structures rarely coincides with the shape of the divergent beam. Consequently, some structures are imaged with more inherent distortion than others, and shapes of anatomic structures can be entirely misrepresented. Structures farther from the IR will be distorted (i.e., magnified) more than those closer to the IR; structures closer to the x-ray source will be  distorted (i.e., magnified) more than those farther from the x-ray source.
  • For the shape of anatomic structures to be accurately recorded, the structures must be parallel to the x-ray tube and the IR, and aligned with the central ray (CR). The shape of anatomic structures lying at an angle within the body or placed away from the CR will be misrepresented on the IR. There are two types of shape distortion. If a linear structure is angled within the body, that is, not parallel with the long axis of the part/body and not parallel to the IR, then that anatomic structure will appear smaller—it will be foreshortened. On the other hand, elongation occurs when the x-ray tube is angled. Image details placed away from the path of the CR will be exposed by more divergent rays, resulting in rotation distortion. This is why the CR must be directed to the part o greatest interest. Unless the edges of a three-dimensional object conform to the shape o the x-ray beam, blur or unsharpness will occur at the partially attenuating edge of the object. This can be accompanied by changes in radiographic/image density, according to the thickness of areas traversed by the x-ray beam
Exit mobile version