Special Hearing Tests
| A | the electrode must be kept close to source of amplitude | |
| B |
actionpotentional is produced by inner hair cells and positive waves |
|
| C |
cochlearmicrophonic is produced from the outer hair cells of the cochlea |
|
| D |
mostly used in monitoring the meniers disease |
All are true about electrocochleography except
| A |
the electrode must be kept close to source of amplitude |
|
| B |
actionpotentional is produced by inner hair cells and positive waves |
|
| C |
cochlearmicrophonic is produced from the outer hair cells of the cochlea |
|
| D |
mostly used in monitoring the meniers disease |
Electrodes can be kept over the promontory,EAC or outside,the signals are better as much it is kept close to source.
Different waves are cochlear microphonic:produced from outer hair cells action potential is produced by all the nerve fibers Used in diagnosis and monitoring the meiners disease and endolymphatic hydrops
Which of the following is the investigation of choice in assessing hearing loss in neonates?
| A |
Impedance audiometry |
|
| B |
BERA |
|
| C |
Free field audiometry |
|
| D |
Behavioral audiometry |
Brainstem Auditory Evoked Responses (BERA)
>The BERA test measures the reaction of the parts of a child’s nervous system that affect hearing. The ABR test measures the auditory nerve’s response to sounds.
>It is performed if a newborn or infant fails the hearing screening test given in the hospital shortly after birth, or for older children if there is a suspicion of hearing loss that was not confirmed through more conventional hearing tests.
Otoacoustic emissions (OAE)
>Otoacoustic Emission (OAE) Screening is used widely in hospital-based newborn hearing screening programs.
>Portable OAE screening is the most practical method for screening infants and toddlers.
Otoacoustic emissions arise from:
| A |
Inner hair cells |
|
| B |
Outer hair cells |
|
| C |
Both inner and outer hair cells |
|
| D |
Organ of Corti |
OTOACOUSTIC EMISSIONS (OAEs):
- They are low-intensity sounds produced by outer hair cells of a normal cochlea and can be elicited by a very sensitive microphone placed in the external ear canal and analyzed by a computer.
- The sound produced by outer hair cells travels in a reverse direction:
- OAEs are present when outer hair cells are healthy and are absent when they are damaged and thus help to test the function of the cochlea.
- They do not disappear in VIIIth nerve pathology as cochlear hair cells are normal.
Types of OAEs:
- spontaneous
- Evoked.
- OAEs are used as a screening test of hearing in neonates and to test hearing in uncooperative or mentally challenged individuals after sedation.
- They help to distinguish cochlear from retro-cochlear hearing loss.
- OAEs are also useful to diagnose retro cochlear pathology, especially auditory neuropathy.
| A |
Otosclerosis |
|
| B |
Presbyaccusis |
|
| C |
Acoustic nerve schwannoma |
|
| D |
Otitis media with effusion |
Recruitment Phenomenon
- It is a phenomenon of abnormal growth of loudness
- The ear which does not hear low-intensity sound begins to hear greater intensity sounds as loud or even louder than a normal hearing ear.
- Thus a loud sound which is tolerable in the normal ear may grow to abnormal levels of loudness in the recruiting ear and thus become intolerable
- Recruitment is typically seen in lesions of cochlea i.e. Meniere’s disease”, presbycusis.
- Patients with recruitment are poor candidates for hearing aids.
| A |
Pure Tone Audiometry |
|
| B |
Stapedial Reflex |
|
| C |
Otoacoustic Emissions |
|
| D |
Brainstem evoked auditory response |
Otoacoustic emissions (OAEs) are objective, noninvasive, and rapid measures (typically less than 2 minutes) used to determine cochlear outer hair cell function.
| A |
Free field audiometry |
|
| B |
Behavioral audiometry |
|
| C |
Impedance audiometry |
|
| D |
Brainstem Evoked Response Audiometry (BERA) |
Brainstem evoked response audiometry is the investigation used to detect acoustic neuromas or determine hearing thresholds in children.
In Brainstem Evoked Response Audiometry electrodes are placed over various areas of the skull and sounds of various frequencies are projected to the ear with a headphone. The potentials generated by the brain in response to the sound are recorded and studied.
Which statement regarding electrocochleography is TRUE?
| A |
A method of recording auditory evoked response taken from cochlea and auditory nerve |
|
| B |
Summation potential (SP) is a component of synchronised auditory nerve potentials |
|
| C |
Cochlear microphonics is resting receptor potential of outer hair cells |
|
| D |
Whole nerve AP represent direct current endocochlear AP produced by an acoustic stimulation |
Electrocochleography is a near-field technique that monitors the most peripheral components of the auditory system. It measures electrical potentials arising in the cochlea and auditory nerve in response to auditory stimuli within first 5 milliseconds.
Auditory stimuli are delivered to the cochlea commonly in the form of a wideband click, composed of many frequencies that stimulate the entire cochlea. Cochlear potentials that can be recorded from humans include the cochlear microphonic potentials, the summating potential(SP) and the compound action potential(AP) of the auditory nerve. The SP and AP are the most important components of electrocochleography.
To find threshold of hearing in young infants and children to within 5-10 dB
| A |
Otoacoustic emission (OAE) |
|
| B |
Auditory brainstem response (ABR) |
|
| C |
Pure tone audiometry (PTA) |
|
| D |
Free field audiometry |
Most screening programmes for newborn hearing ability uses otoacoustic emission as the initial test. It this test fails then auditory brainstem response is used for screening.
Otoacoustic emissions are low level sound emitted spontaneously by the cochlea on presentation of an auditory stimulus. Persons with normal hearing produce oatoacoustic emissions whereas those who have hearing loss of 30-40db HL or greater do not produce OAE.
Auditory brainstem response testing is used with babies between the ages of birth and 5 months. It is the electrophysiological response to an acoustic stimulus and originates from the eight cranial nerve and auditory brain stem. It can be used to determine the degree of hearing loss at different audiometric frequencies.
Auditory screening is required in children under following conditions –
| A |
Before doing exchage transfusion for hyperbilirubinemia |
|
| B |
In premature babies |
|
| C |
Before starting aminoglycoside therapy |
|
| D |
All |
Ans. is ‘a’ i.e., Before doing exchange transfusion for hyperbilirubinemia, ‘b’ In premature babies,`c’ i.e., Before starting aminoglycoside therapy
o All neonates with risk factor for hearing loss must be screened with an auditory brainsteam response (ABR) or otoacoustic emissions (OAE) test to exclude hearing impairment.
Indications for Auditory screening ‑
Family history or hereditary childhood SNHL.
- In utero infections – Toxoplasmosis, Rubella, CMV, herpes, syphilis.
- Craniofacial anomaly-morphological abnormalities of the pinna and ear canal.
- Birth weight less than 1500 gm.
- Hyperbilirubinemia at a serum level requiring exchange transfusion.
- Ototoxic drugs aminoglycosides, loop diuretics
- Bacterial meningitis
- Apgar score 0 to 4 at 1 min or 0 to 6 at 5 min.
- Mechanical ventilation lasting more than 3 days.
Stigmata associated with a syndrome known to include hearing loss.
| A |
Are by product of outer hair cell |
|
| B |
Are by product of inner hair cell |
|
| C |
Used as a screening test of hearing in newborn infant |
|
| D |
a and c |
To distinguish between cochlear and post cochlear damage, test done is:
| A |
Brainsterm evoked response audiometry |
|
| B |
Impedence audiometry |
|
| C |
Pure tone audiometry |
|
| D |
Auditory cochlear potential |
BERA is very useful in distinguishing between cochlear pathology and retrocochlear pathology for SNHL
Impedence audiometry and PTA tests the middle ear pathology.
In electrocochleography:
| A |
It measures middle ear latency |
|
| B |
Outer hair cells are mainly responsible for cochlear microphonics and summation potential |
|
| C |
Summation potential is a compound of synchronus auditory nerve potential |
|
| D |
Total AP represents endocochlear receptor potential to an external auditory stimulus |
Ans. is. B. Outer hair cells are mainly responsible for cochlear microphonics and summation potential
Which of the following is true about meniere’s disease?
| A |
Surgery is the mainstay of treatment |
|
| B |
Electrocochleography is the gold standard investigation for diagnosis |
|
| C |
Semont’s maneuver is used for treatment |
|
| D |
In initial stages, inverted ‘V’ shaped audiogram is seen |
Electrocochleography is diagnostic in Meniere’s disease.
Extra-tympanic ECochG is a non-invasive technique to record the amplitude of summating potential (SP) and (AP) action potential amplitute alone or the amplitude ratio of SP and action potential (AP)….(SP/AP)
Ratio of Summating Potential (SP) to Action Potential (AP) is greater then 30%. in patients with MD.
The screening investigation of high risk neonates in ICU for suspected hearing loss is:
| A | Otoacoustic emissions | |
| B |
Free field audiometry |
|
| C |
Stapedial reflex testing |
|
| D |
Pure tone audiometry |
The screening investigation of choice of high-risk neonates in ICU for suspected hearing loss is otoacoustic emissions.
Otoacoustic Emissions Test (OAEs)
A miniature earphone and microphone are placed in the ear, sounds are played and a response is measured.
- If a baby hears normally, an echo is reflected back into the ear canal and is measured by the microphone
- When a baby has a hearing loss, no echo or a reduced echo can be measured on the OAE test
| A |
Retrocochlear lesions |
|
| B |
Otosclerosis |
|
| C |
Meniere’s disease |
|
| D |
None of the above |
Recruitment was defined as an abnormal growth of response with increasing stimulus intensity.
The theory of recruitment is that as the hair cells in your cochlea become ineffective, they “recruit” their (still working) neighbor hair cells to “hear” the frequency the damaged hair cell was supposed to hear, in addition to the frequency the still working hair cell was supposed to hear. This increases the signal from the still working hair cells.
Best time for hearing assessment in an infant ‑
| A |
1st month of life |
|
| B |
3-6 months |
|
| C |
6-9 months |
|
| D |
9-12 months |
Ans. is ‘a’ i.e., 1st month of life
The American Academy of Pediatrics (AAP), Joint Committee on Infant hearing (2007), has recommended that all newborn infants be screened for hearing impairment either as neonate or before 1 month of age and that those infants who fail newborn screening have an audiologic examination to varify hearing loss before age of 3 months.